ABSTRACT As the main detector of the Gravitational wave high-energy Electromagnetic Counterpart All-sky Monitor mission (GECAM), the calibration of the energy response and detection efficiency of the gamma-ray detector (GRD) is the main content of the ground-based calibration. This article mainly focuses on the calibration of the energy response and detection efficiency in the 8–160 keV with a refined measurement around the absorption edges of the lanthanum bromide crystal ($\rm {LaBr_3}$). The GRD performances for different crystal types, data acquisition modes, working modes, and incident positions are also analysed in detail. We show that the calibration campaign is comprehensive, the detector performance meets the flight requirements, and the calibration results generally agree with simulations as expected. The detector’s model was corrected by the ground-based calibration, which led to the establishment of the calibration data base.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.