Abstract

The reliability of CT size and density measurements has been examined by using phantom measurements with constant examination and reconstruction parameters (Somatom DRH/Siemens). Ten plexiglass rods with diameters from 2 to 15 mm. were measured singly in (1) air, (2) isodense (146 Hu) and (3) hyperdense medium (985 Hu) using 2 mm. sections. In the isodense medium there was only small variation in density measurements (124 to 131 Hu). Measurements of size showed a maximum error up to -35%. Density measurements in the hyperdense medium produced raised values (197 +/- 17 Hu at 15 mm.). There was an even greater increase for rods smaller than 5 mm. Measurement of size was correct, as it was in air (maximum deviation 8%). Density measurements in air for diameters less than 6 mm. were significantly reduced.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.