Abstract
This work presents a straightforward, highly sensitive, and cost-effective method for the simultaneous determination of V, Ti, Ni and Ga by high resolution-continuum source electrothermal atomic absorption spectrometer (HR-CS ETAAS) in aqueous environmental samples (tap and seawater samples). The system is based on retention of the analyte onto a novel magnetic nanomaterial (M@GO magnetic graphene oxide) functionalised with methylthiosalicilate (MTS). The formed complexes between the M@GO-MTS and the target analytes were broken, adding 1 mL of nitric acid (6%) and sonication for 5 min. The optimized method achieved detection limits of 0.71 μg L-1 for Ti, 0.20 μg L-1 for V, 0.04 μg L-1 for Ga, 0.66 μg L-1 for Ni. The accuracy of the proposed method was demonstrated by analysing two certified reference materials and by determining the analyte content in spiked environmental water samples. The results obtained using this method were in good agreement with the certified values of the standard reference materials, and the recoveries for the spiked tap water and seawater samples ranged from 94% to 120%.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Analytical methods : advancing methods and applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.