Abstract
Shifts in the displayed shades of gray were found in the initial evaluation of a gray scale ultrasound display. Quantitative measurements of these shifts were performed. These measurements, along with measurements of scan converter output and photographic exposure time, established television (TV) performance as the major cause of shifts. Random variations in photographic exposure time were shown to cause small random shifts. Recommendations are given to minimize the effects of these shifts in clinical sonograms. To minimize shifts, hardware modifications were made on a similar TV. These modifications substantially improved the TV performance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.