Abstract

The design of grating monochromators for the spectral selection of extreme-ultraviolet femtosecond pulses without altering significantly the pulse time duration is discussed here. Time-preserving monochromators are realized in a single-grating configuration if the number of illuminated grooves is minimum for a given spectral resolution; therefore the grating time response is close to the Fourier limit for such a resolution. The classical-diffraction mount and the off-plane mount are compared for applications in the extreme-ultraviolet region—the classical mount should be preferred for monochromators with time responses in the 100–200 fs range, while the off-plane mount gives responses in the 10–50 fs range, and is therefore suitable to realize femtosecond time-preserving monochromators.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.