Abstract

A bare graphite-polyurethane composite was evaluated as an amperometric detector in the flow injection determination of atenolol in pharmaceutical formulations. Using a flow injection analysis (FIA) procedure, a linear analytical curve was observed in the 0.2–3.0 mmol L−1 range with a minimum detectable net concentration limit of detection (LOD = 18.1 µmol L−1 and 90 determinations h−1. Interferences from propranolol and furosemide were observed but not from the other components of the tablet. Thus, it was possible to determine atenolol in tablets without interference and with results that agreed with high performance liquid chromatography (HPLC) procedure with a 95% confidence level in a fast and accurate procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.