Abstract

A graphical procedure was proposed for the optimum design of transdermal drug-delivery systems enhanced by iontophoresis. Contour plots displayed the relationships among steady-state plasma level, current density and initial drug concentration in a vehicle. This information was combined with a closed-form expression of the process time constant, estimated as the medicament in the blood reaches a plateau after application of the electric field. Analysis was conducted using Laplace-transformed variables and did not require time-domain solutions. Simulation results show that a current density of 0.044mA/cm2 and a loading of 3500μg/ml of dexamethasone sodium m-sulfobenzoate were necessary to achieve an equilibrium plasma concentration of 1.254ng/cm3 with a time constant of 8.34h.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.