Abstract
To preliminarily investigate a technical feasibility of a submillisecond echotime concurrent-dephasing-and-excitation (CODE) sequence for pulmonary MRI on clinical and preclinical MR scanners Methods: CODE imaging (echo time, 0.14 ~ 0.18 ms) was performed with American College of Radiology phantom at 3T, 7 healthy volunteers at 1.5 and 3T, 10 rabbits at 3T, and 2 rodents at 9.4T. Signal-to-noise ratio was compared in phantom. Image quality of human MRI was visually assessed on a 5-point scale for comparison between CODE and conventional lung MRI sequences. Visibility of bronchi, subcentimeter nodules, and MR air-bronchogram were assessed in animal studies. In phantom study, signal-to-noise ratio was higher with CODE than with original three-dimensional ultrashort-echotime sequence (106.71 ± 4.32 vs 91.66 ± 3.54; p < 0.001). Image quality of human MRI was better with CODE than with conventional MRI sequences (p ≤ 0.002). Bronchi remained traceable up to the fifth bronchial generation in CODE images in rabbits and rodents. 95.2% of metastatic nodules (diameter, 1.5 ±0.4mm) and 93.8% of MR air-bronchogram (diameter, 0.9 ±0.2 mm) in rabbits. Submillisecond echotime pulmonary MRI was technically feasible by using CODE on various MR scanners. Advances in knowledge: CODE can be a practical alternative for lung MRI on both clinical and pre-clinical scanners, without challenges of free-induction-decay-based ultrashort-echo time sequences.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.