Abstract
Gonioreflectometric determination of reflectance factors that involves hemispherical collection of reflected flux, which is an alternative to integrating sphere-based methods, is discussed. A detailed description of a gonioreflectometer built at the Helsinki University of Technology is presented. The instrument is used to establish an absolute scale of total diffuse reflectance factors throughout the spectral range 360-830 nm. The hemispherical reflectance factors are obtained through integration of the gonioreflectometric measurement results. The reflectance factors of white high-quality artifacts can be determined with a combined standard uncertainty of 0.20%. Results of test measurements were found to be in agreement with values traceable to other absolute scales based on integrating-sphere methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.