Abstract

To assess the validity of the Tono-Pen XL as an alternative to the Goldmann applanation tonometer (GAT) for the measurement of intraocular pressure (IOP) in patients with ocular hypertension (OHT) and glaucoma. Over a 3 month period, patients with OHT or glaucoma attending a general clinic had IOP measurements taken using the Tono-Pen XL and the GAT. Tono-Pen measurements were taken by suitably trained nursing staff, while Goldmann tonometry was performed by the examining ophthalmologist. There were 124 eyes of 62 patients in this study. Overall, mean IOP was 18.3 +/- 4.8 mmHg using GAT and 18.8 +/- 5.5 mmHg using the Tono-Pen. Using the Bland-Altman method, the upper and lower limits of agreement between the two devices were +10.6 mmHg and -9.6 mmHg, respectively. Significant over- and under-estimates of IOP were noted in 10 (16%) patients. Our findings suggest that the Tono-Pen XL cannot be used as a substitute for GAT in the management of patients with glaucoma or OHT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.