Abstract

ABSTRACT.Pattern electroretinograms (PERGs) were recorded in ten subjects employing two varieties of electrodes: a gold foil electrode inserted into the lower fornix, and a gold cup skin electrode overlying the inferior orbit. Recordings were obtained as subjects viewed a checker-board pattern that reversed with a frequency of 2.1 Hz. Check size varied from 4 degrees to 0.25 degrees, at a contrast of 98%. Two additional recordings were obtained with 0.5 degree check size, lower luminance, and lower contrast. The PERG recorded with a skin electrode was similar in configuration to the PERG recorded with a gold foil electrode, but smaller in amplitude. The equation relating the amplitude of the b-wave recorded from the foil electrode (y) to the amplitude of the b-wave recorded from the skin electrode (x) was: y = 0.127 + 2.150x, with an R value of 0.82. Latencies of both the a- and b-waves were comparable under both recording conditions. We conclude PERGs may be adequately recorded with a noncorneal skin ele...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.