Abstract

AimsTo describe Hybrid closed-loop (HCL) and advanced hybrid closed-loop (AHCL) performance in the hospital setting based on the continuous glucose monitoring (CGM) metrics description. MethodsThis was an observational study from a cohort of patients with T1D using HCL/AHCL with history of hospitalization. CGM metrics were analyzed during the hospital stay. CGM metrics sub-analysis of the population with active Automated Mode (AM) and SmartGuard (SG) during hospitalization and/or surgical procedure was performed. ResultsTwenty-four patients were included (50 % women; mean age, 49 years [inter-quartile range (IQR), 39–62 years]). During hospitalization 70.8 % patients achieved %Time in Range (TIR) between 70 and 180 mg/dL ≥ 70 %. The overall %TIR was 75.5 % (IQR, 67.3–81.5 %), % time below range (TBR) < 70 mg/dL was 2.1 % (IQR, 0.7–5.4 %) and %TBR < 54 mg/dL was 0 % (IQR, 0–5.4 %). Users of the AHCL with active SG achieved a non-significant higher %TIR during hospitalization (79 % [73.8.88 %] vs. 76 % [72.81 %], p = 0.312) and had a shorter stay (3[IQR, 2.4] vs. 6 days[IQR, 5.7], p = 0.045) compared to the users of the HCL with AM active. No device-related serious adverse events occurred for users of either system. ConclusionsHCL/AHCL systems with active AM/SG in patients with T1D in the hospital environment leads to %TIR > 70 % in ranges of 70–180 mg/dL in patients without increasing hypoglycemia.