Abstract

The second generation of a competitive ELISA for prolamin quantification based on the R5 antibody was studied for method performance and suitability to detect partially hydrolyzed prolamins in food. To be able to convert signal intensities to gluten concentrations, as required by the Codex Alimentarius Standard, a new calibrator consisting of a peptic-tryptic digest of wheat, rye, and barley prolamins was used for the first time. LOD and LOQ of the assay were 1.36 and 5.0 mg prolamin/kg food, respectively. Analysis of beer samples and a hydrolyzed wheat product showed that the assay provided significantly higher prolamin concentrations, compared to the sandwich ELISA based on the same antibody, which is only suitable for the detection of intact prolamins. Spiking experiments with defined concentrations of partially hydrolyzed prolamins gave recoveries ranging from 92 to 136%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.