Abstract

The causes of porcine valve failure were evaluated pathologically in 129 bioprostheses (BP) removed at surgery or necropsy. Fifty-one BP had been in place for less than two months (early explants); most of the complications of this group were observed in the mitral position: left ventricular outflow obstruction (43 percent), thrombosis (23.5 percent), myocardial dysruption (20 percent), cardiac rupture (10 percent), and perivalvular leak (3.5 percent). Among the 78 explants in place for more than two months (late explants), 66 had evidence of dysfunction. Calcification was the leading cause of failure (46 percent), followed by thrombosis (17 percent), endocarditis (12 percent), fibrous tissue overgrowth (10 percent), perivalvular leak (9 percent), and primary cusp tears (6 percent). Calcific degeneration is a hallmark of the phenomenon of biodegradation, and experimental animal models are encouraged since they may represent the clue to preventing calcification and therefore enhancing performance and durability of the porcine BP.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.