Abstract

An optical ring resonator based sensor is proposed for bio-sensing applications. It is applied to glucose level monitoring in blood samples in the range 10 to 200 mg/dl using minimal invasive technique. Two configurations of ring resonator are proposed that are based on ridge and slot waveguide. The average sensitivity of slot ring resonator is observed to be 360 nm/RIU, which is six times higher (60 nm/RIU) than that of ridge ring resonator. Both the sensor configurations are adept to respond to slight variation in glucose concentration, which affects the refractive index and ultimately the spectral shift of the resonator.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.