Abstract

The superluminescent diode based differential absorption frequency domain low-coherence interferometry (FD-DALCI) technique is proposed and demonstrated for sensing physiological concentrations of glucose (0-250mg/dl) in oral mucosa simulating phantoms (intralipid of concentrations 0.25-0.50%) with wavelengths at 1589 and 1310nm. The proposed technique allows simultaneous measurements of refractive index based spectral shift and estimation of physiological concentration of glucose in intralipid with scattering characteristics using the differential absorption approach. The sensitivity of the glucose concentration obtained by spectral shift measurement was ≈0.016 nm/(mg/dl), irrespective of the intralipid concentration. The resolution of the glucose level was estimated to be ≈15 mg/dl in 0.25% intralipid and ≈19 mg/dl in 0.5% intralipid using the FD-DALCI technique.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.