Abstract

Closed-loop diabetes management has been shown to indicate improved glycaemic control and better compliance over open loop diabetes management. Currently, commercially available diabetes management devices rely on continuous glucose monitoring (CGM) sensors which monitor glucose levels from the interstitial fluid (ISF). As there exists a physiological delay between the blood glucose levels compared to the ISF glucose levels, it is crucial to predict or forecast glucose levels, in order to prevent hyperglycaemic events due to delayed insulin dosing. Accuracy of the forecast also needs to be optimum such that overdosing on insulin does not lead to hypoglycaemia. In this paper, we describe a novel Long Short Term Memory (LSTM) network which follows a wide and deep approach for different features to deliver an accurate glucose prediction output. It achieved a Mean Absolute Relative Difference (MARD) of 2.61 and Root Mean Squared Error (MSE) of 5.04. Clinical relevance- This work is relevant for closed-loop diabetes management devices, which are currently being used to manage Type 1 Diabetes (T1D).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.