Abstract
OBJECTIVEThe current standard for hospital glucose management is point-of-care (POC) testing. We conducted a randomized controlled trial of real-time continuous glucose monitoring (RT-CGM) compared with POC in a non–intensive care unit (ICU) hospital setting.RESEARCH DESIGN AND METHODSA total of 110 adults with type 2 diabetes on a non-ICU floor received RT-CGM with Dexcom G6 versus usual care (UC). RT-CGM data were wirelessly transmitted from the bedside. Hospital telemetry monitored RT-CGM data and notified bedside nursing of glucose alerts and trends. Standardized protocols were used for interventions.RESULTSThe RT-CGM group demonstrated significantly lower mean glucose (M∆ = −18.5 mg/dL) and percentage of time in hyperglycemia >250 mg/dL (−11.41%) and higher time in range 70–250 mg/dL (+11.26%) compared with UC (P values <0.05). Percentage of time in hypoglycemia was very low.CONCLUSIONSRT-CGM can be used successfully in community-based hospital non-ICU settings to improve glucose management. Continuously streaming glucose readings may truly be the fifth vital sign.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.