Abstract

The COVID-19 pandemic has increased demand for point-of-care (POC) screening tests such as lateral flow assays (LFAs) and highlighted the need for sensitive and cost-effective POC diagnostic platforms. Here, we demonstrate an LFA platform using standard fluorescent nanoparticle reporters in which optical excitation is replaced by chemical excitation using the peroxyoxalate-based chemistry of inexpensive, shelf-stable glowsticks. The one-step chemi-excitation of fluorescent particles produces visible light readable by an unmodified smartphone, enhancing sensitivity while preserving simplicity and cost-effectiveness. Our Glow LFA detected the common model analyte human chorionic gonadotropin with a limit of detection (LoD) of 39 pg/mL—over ten times more sensitive than standard gold nanoparticles using the same antibodies. We also demonstrate its application to the detection of SARS-CoV-2 nucleoprotein at 100 pg/mL in nasal swab extract. Multiple fluorescent dyes can be chemi-excited by a single reagent, allowing for color multiplexing on a single LFA strip with a smartphone camera. The detection of three analytes on a single LFA test line was demonstrated using red, green, and blue fluorescent reporter particles, making glow LFA a promising platform for multiplexed detection.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call