Abstract
The influence of the maximum readout temperature on the LiF GR-200 dose response was studied. A fast heating rate (8°C.s -1 ) and a new method to set the maximum readout temperature were employed for this study. The new method determines on each measurement the maximum temperature with reference to the position of the main TL peak. This is achieved through the on-line analysis of the glow curve during heating. The good measurement reproducibility attained by using this method and a previously described computerised method for TL evaluation is also presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.