Abstract

The global thrombosis test (GTT) is a point of care device that tests thrombotic and thrombolytic status. The device exposes whole blood flow to a combination of both high and low shear stress past and between ball bearings potentially causing thrombin and fibrin formation. The question arises as to whether thrombosis in the GTT is dominated by coagulation-triggered red clot or high shear-induced white clot. We investigated the nature of the thrombus formed in the GTT, the device efficacy, human factors use, and limitations. The GTT formed clots that were histologically fibrin-rich with trapped red blood cells. The occlusion time (OT) was more consistent with coagulation than high shear white clot and was strongly lengthened by heparin and citrate, two common anticoagulants. The clot was lysed by tissue plasminogen activator (tPA), also consistent with a fibrin-rich red clot. Changing the bead to a collagen-coated surface and eliminating the low shear zone between the beads induced a rapid OT consistent with a platelet-rich thrombus that was relatively resistant to heparin or tPA. The evidence points to the GTT as occluding primarily due to fibrin-rich red clot from coagulation rather than high shear platelet aggregation and occlusion associated with arterial thrombosis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.