Abstract

Near-infrared (NIR) global quantitative models were evaluated for the moisture content (MC) determination of three different freeze-dried drug products. The quantitative models were based on 3822 spectra measured on two identical spectrometers to include variability. The MC, measured with the reference Karl Fischer (KF) method, were ranged from 0.05% to 4.96%. Linear and non-linear regression models using Partial Least Square (PLS), Decision Tree (DT), Bayesian Ridge Regression (Bayes-RR), K-Nearest Neighbors (KNN), and Support Vector Regression (SVR) algorithms were created and evaluated. Among them, the SVR model was retained for a global application. The Standard Error of Calibration (SEC) and the Standard Error of Prediction (SEP) were respectively 0.12% and 0.15%. This model was then evaluated in terms of total error and risk-based assessment, linearity, and accuracy. It was observed that MC can be fastly and simultaneously determined in freeze-dried pharmaceutical products thanks to a global NIR model created with different medicines. This innovative approach allows to speed up the validation time and the in-lab release analyses.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call