Abstract

For more than 30 years ceramics are used for heads and acetabular cups of hip joint arthroplasty. The results of wear rates of metal-on-polyethylene, ceramic-on-polyethylene and ceramic-on-ceramic determined from a hip simulator testing were not confirmed when comparing to results of clinical follow ups using either X-rays methods or investigating retrieved implants. There is consensus that the wear couple of metal-on-polyethylene is twice or three times larger than the one for ceramic-on-poly-ethylene. Ceramic-on-ceramic offers the option of 5μm/year. When evaluation novel concepts the best is to investigate retrievals because the hip simulator does not mimic all in-vivo effects. Hip simulator testing is not worse case testing as it should be, hip simulator testing is best case testing. The results may be used for screening purposes.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call