Abstract
This paper demonstrates the feasibility of using glass rod detectors for quality assurance audit of radiosurgery units. Five radiosurgery units (3 Gamma Knife model C, 1 Gamma Knife model U and 1 Cyberknife) located in California participated in the study. At each center glass rod detectors were used to measure a number of dosimetric parameters including relative collimator output factor and absolute dose rate. The Gamma Knife data obtained is in excellent agreement with the commissioning data generated by the manufacturer for each unit and the Cyberknife data is in general agreement with the data published by other centers. In particular the output factor of the 4 mm Gamma Knife helmet factor, a subject of abundant debate, was measured in the range 0.863-0.872 with an accuracy of better than 1% across the four participating centers. It is hoped that this pilot study will facilitate a nationwide postal audit of stereotactic radiosurgery units.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.