Abstract
This paper describes the construction and evaluation of a Ka band bolometer receiver using a highly temperature-resistance dependent (at 4.2°K) doped germanium bolometer of Texas Instruments, Inc. The receiver noise equivalent power (NEP) of approximately 10−9 W and input bandwidth exceeding 12 Gc has a rms signal level indicating uncertainty equivalent to 90 C° per cycle of output filtering. Evaluation of the germanium bolometer as a millimeter wavelength radiometer detector was performed at Ka band to utilize existing apparatus and to facilitate fabrication of the receiver. With the increased sensitivity anticipated for smaller bolometer elements, short millimeter and submillimeter wavelength scaled models of the receiver appear promising as detectors for radio telescopes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.