Abstract

A major challenge facing the development of portable, low-cost NMR is the development of robust yet sensitive transceivers, for which several trade-offs in scalability, performance and complexity are usually necessary. Here we report on a stripline-based NMR detector that overcomes previous limitations. It features a sensitivity of 5.7×10-4 TA−1Ω−0.5 over a sample volume of 10 mm × 10 mm × 3 mm, an exceptionally high B1 homogeneity of A450/A90=98.4%, intrinsic electromagnetic shielding of 27 dB from environmental influences, and a total signal gain of 68 dB in the presence of a noise factor of 1.28, without any exterior shielding. The new dual-coil arrangement offers a downscalable geometry optimised for gap magnets, and it is voltage-tunable and plug-in compatible with commercial software-defined radio spectrometer boards. Exceptionally, it features both common-mode and novel differential-mode NMR measurement abilities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.