Abstract

A monopivot magnetic suspension blood pump has been developed in our laboratory. The flow patterns within the pump should be carefully examined in order to prevent thrombogenesis, especially around the pivot bearing. Therefore, the effects of the pump geometry on the local flow were analyzed using computational fluid dynamics together with the experimental flow visualization. The engineering goal was to reduce the area of stagnation around the pivot in order to prevent thrombus formation. As a result, the stagnation area and the flow rate through the washout holes were found to be highly affected by the size and geometry of the washout holes. Secondary flow was revealed to form a jet-like wash against the pivot, thus preventing thrombus formation. The flow rate through the washout holes was estimated to be up to one fifth of the pump flow rate, depending on the cross-sectional areas of the washout holes. Furthermore, an anti-thrombogenic effect was attained by removing a small gap between the male and female pivots.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.