Abstract

A geometric method of measuring absolute left ventricular volumes from gated blood-pool studies obtained in a single-plane modified left anterior oblique view was evaluated prospectively in 30 patients who also underwent single-plane contrast ventriculography. The gated studies used a 30 degrees straight-bore slant-hole collimator with the holes slanted caudally. Left ventricular end-diastolic volume was calculated using the area-length method, with semiautomatic definition of the left ventricular region of interest and the maximum length of the left ventricle. Ejection fraction was determined from the left ventricular time/activity curve. The left ventricular end-systolic volume was derived using the end-diastolic volume and ejection fraction. Correlation coefficients between the two methods were 0.93 for end-diastolic volume, 0.95 for end-systolic volume, and 0.91 for ejection fraction. This method provided accurate and highly reproducible measurements of actual left ventricular volumes and was easily applicable in routine clinical studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.