Abstract

Patient immobilization is an important element for the geometric accuracy of stereotactic radiotherapy (SRT) techniques. However, differences in the immobilization and accuracy with the use of a G frame versus a double remain unknown. The aim of this retrospective comparative cross-sectional study was to compare the extents of geometrical shifting on verification of stereotactic techniques with the use of a G frame versus a double mask. The study cohort included patients who received RT between February and March 2016 at the Department of Radiotherapy of Cipto Mangunkusumo Hospital. The t-test was used to compare geometrical shifting values on verification with the use of a G frame versus a double mask. The mean value of geometrical shifting in patients with SRT with the use of a double mask on the laterolateral, craniocaudal, and anteroposterior axes were 0.4 ± 0.3, 0.5 ± 0.4, and 0.5 ± 0.4 mm, respectively, while values with the use of a G frame were 0.3 ± 0.2, 0.3 ± 0.4, and 0.4 ± 0.3 mm, respectively. There were no differences in the mean geometrical shifting values with the use of a double mask versus a G frame.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.