Abstract

Generally, three devices (the Genucom Knee Analysis System, the MEDmetric KT-1000 Arthrometer, and the Stryker Knee Laxity Tester), which have been presented in the literature, may be used to assess clinically the amount of knee laxity in an objective manner. This study compared the reproducibility of the anterior and posterior knee laxity values for each and made direct comparisons of the results obtained. Thirty asymptomatic subjects were given a test-retest protocol on both legs with all three devices. Devices were tested in a counterbalanced order. Repeatability of test values within devices was variable, but all were acceptable, the lowest being r = 0.74. Analysis of variance (AN-OVA), and correlational analysis revealed that device-specific anterior and posterior laxity values were produced. We concluded that each commercially available knee laxity testing device can provide reproducible quantitative measurements of knee laxity; however, due to differences in device sensitivities and functional design, numerical results from one device cannot be generalized to another device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.