Abstract

Characterizing polymerase chain reaction (PCR) amplicons has been accomplished for the first time using flow injection analysis coupled to electrospray ionization mass spectrometry (ESI-MS). The PCR amplicons were amplified at the human tyrosine hydroxylase short tandem repeat locus from an individual homozygotic for the 9.3 allele. One product was amplified using Pfu polymerase and yielded a blunt-ended amplicon of 82 base-pairs (bp) in length. The second PCR product was amplified using Taq polymerase that resulted in an amplicon with cohesive termini of 82 bp plus either mono- or diadenylation. The two PCR amplicons were alternatively injected using a 0.5-microL loop at 2 microM for the Pfu amplicon and 1 microM for the Taq amplicon with a flow rate of 200 nL/min during data acquisition. Both PCR amplicons were accurately identified using mass measurements illustrating the compatibility of ESI-MS for genotyping short tandem repeat sequences and the potential for high-throughput genotyping of large PCR amplicons.

Full Text
Published version (Free)