Abstract

Generic dry-contact ear-EEG allows for discreet, user-friendly, unobtrusive, cost-effective and convenient recordings of EEG in real-life settings. In this study we introduce a new generic earpiece design with larger internal ear electrode distances, resulting in an increased spatial coverage compared to previous generic earpiece designs. The signal quality of ear-Fpz, within-ear (the measuring and reference electrode located in the same ear) and cross-ear (the measuring electrodes located in one ear and the reference electrode in the opposite ear) electrode configurations of the developed generic earpiece was evaluated with auditory steady-state responses (ASSR) and compared to dry-contact cEEGrid. Ten subjects with different ear sizes were included. The recordings were performed in a sleep setup, where the subjects were lying on a bed and the effect of sleeping position (back vs. sides) was investigated. We found that the generic earpiece attained statistically significant ASSRs with ear-Fpz, within-ear and cross-ear electrode configurations. However, the dry-contact cEEGrid achieved significantly higher average ASSR signal-to-noise ratio (SNR) compared to the generic earpiece. Additionally, this study showed no significant difference between back and side positions for the ear-EEG.