Abstract
Aiming to develop the compact nuclear magnetic resonance devices, the authors have been developing the uniform magnetic fields in the space between the face-to-face magnetic poles which contain HTS bulk magnets. The authors deformed the original convex shape of the magnetic field distribution to concave by attaching a ferromagnetic iron plate of 20 mm and 100 mm in diameter on one of the pole surfaces. The magnetic field distribution precisely measured along x-axis changed from concave to convex with increasing distance from the magnetic pole surface, which suggested the presence of magnetically flat region. From the results of mapping on the field uniformity data, we have succeeded in obtaining the field uniformity under 500ppm in the space of 4 x 4 x 0.8 mm3 when the magnetic poles emitting concave and convex field distributions were settled face-to-face with various gaps 30-70 mm. This performance exceeded the target value of 1,500ppm which is necessary to detect the NMR signals.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.