Abstract

BackgroundIt is not unusual to see some parts of tissues are excluded in the field of view of CT simulation images. A typical mitigation is to avoid beams entering the missing body parts at the cost of sub-optimal planning.MethodsThis study is to solve the problem by developing 3 methods, (1) deep learning (DL) mechanism for missing tissue generation, (2) using patient body outline (PBO) based on surface imaging, and (3) hybrid method combining DL and PBO. The DL model was built upon a Globally and Locally Consistent Image Completion to learn features by Convolutional Neural Networks-based inpainting, based on Generative Adversarial Network. The database used comprised 10,005 CT training slices of 322 lung cancer patients and 166 CT evaluation test slices of 15 patients. CT images were from the publicly available database of the Cancer Imaging Archive. Since existing data were used PBOs were acquired from the CT images. For evaluation, Structural Similarity Index Metric (SSIM), Root Mean Square Error (RMSE) and Peak signal-to-noise ratio (PSNR) were evaluated. For dosimetric validation, dynamic conformal arc plans were made with the ground truth images and images generated by the proposed method. Gamma analysis was conducted at relatively strict criteria of 1%/1 mm (dose difference/distance to agreement) and 2%/2 mm under three dose thresholds of 1%, 10% and 50% of the maximum dose in the plans made on the ground truth image sets.ResultsThe average SSIM in generation part only was 0.06 at epoch 100 but reached 0.86 at epoch 1500. Accordingly, the average SSIM in the whole image also improved from 0.86 to 0.97. At epoch 1500, the average values of RMSE and PSNR in the whole image were 7.4 and 30.9, respectively. Gamma analysis showed excellent agreement with the hybrid method (equal to or higher than 96.6% of the mean of pass rates for all scenarios).ConclusionsIt was first demonstrated that missing tissues in simulation imaging could be generated with high similarity, and dosimetric limitation could be overcome. The benefit of this study can be significantly enlarged when MR-only simulation is considered.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call