Abstract

Conformal radiation therapy can be delivered using several methods: intensity-modulated radiotherapy (IMRT) at fixed gantry angles, through the continuous gantry rotation of linac (rotational arc therapy), or by a dedicated treatment unit such as tomotherapy. The recently developed volumetric modulated arc therapy (VMAT), a form of rotational arc therapy, has attracted lots of attention from investigators to explore its capability of generating highly conformal dose to the target. The main advanced features of VMAT are the variable dose rate and gantry rotation speed. In this paper, we present a theoretical framework of generating arbitrary one-dimensional dose profiles using rotational arc therapy to further explore the new degree of freedom of the VMAT technique. This framework was applied to design a novel technique for total body irradiation (TBI) treatment, where the desired dose distribution can be simplified by a one-dimensional profile. The technique was validated using simulations and experimental measurements. The preliminary results demonstrated that the new TBI technique using either dynamic MLC only, variable dose rate only, or a combination of dynamic MLC and variable dose rate can achieve arbitrary dose distribution in one dimension, such as uniform dose to target and lower dose to critical organ. This technique does not require the use of customized compensators, nor large treatment rooms as in the conventional extended SSD technique.