Abstract

Abstract AC magnetic field analyzers with a triaxial coil probe are widely used by health and safety professionals, in manufacturing, and in service industries. For traceable calibration of these analyzers, it is important to be able to generate a stable, homogeneous reference AC magnetic flux density (MFD). In this paper, the generating of AC amplitude MFD value of 150 μT by single-layer Helmholtz type solenoid, described in previous work, was expanded up to a frequency of 100 kHz using the effect of serial resonance. A programmable capacitor array has been developed with a range of adjustable values from 50 pF to 51225 pF. In addition, the multi-layer search coil with a nominal area turns value of 1.3m2, used for adjusting AC MFD in the solenoid, has been modified by a transimpedance amplifier for use in a wider frequency range than up to 3 kHz. The possibility of using the programmable capacitor array up to 150 kHz has also been tested. An AC amplitude MFD value of 150 μT can be generated with expanded uncertainty better than 0.6% up to 100 kHz.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.