Abstract

The test method of Mielenz et al. used to determine the relative stray radiant energy (SRE) level in spectrophotometers is generalized for all sample cell to reference cell thickness ratios greater than unity. It is extended further to include situations where the required ‘cut-off’ solution is not transparent to the SRE. The experimental SRE value which ensued by applying the above method to a Shimadzu 260 ultraviolet/visible spectrophotometer is reconciled with the corresponding experimental values arising from two other SRE test methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.