Abstract
A generalized sidelobe canceler (GSC) approach is proposed for medical ultrasound imaging. The approach uses a set of adaptive weights instead of traditional non-adaptive weights, thus suppressing the interference and noise signal of echo data. In order to verify the validity of the proposed approach, Field II is applied to obtain the echo data of synthetic aperture (SA) for 13 scattering points and circular cysts. The performance of GSC is compared with SA using boxcar weights and Hamming weights, and is quantified by the full width at half maximum (FWHM) and peak signal-to-noise ratio (PSNR). Imaging of scattering point utilizing SA, SA (hamming), GSC provides FWHMs of 1.13411, 1.68910, 0.36195 mm and PSNRs of 60.65, 57.51, 66.72 dB, respectively. The simulation results of circular cyst also show that GSC can perform better lateral resolution than non-adaptive beamformers. Finally, an experiment is conducted on the basis of actual echo data of an ultrasound system, the imaging result after SA, SA (hamming), GSC provides PWHMs of 2.55778, 3.66776, 1.01346 mm at z = 75.6 mm, and 2.65430, 3.76428, 1.27889 mm at z = 77.3 mm, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.