Abstract

We study the possibility of suppressing light shifts in Ramsey spectroscopy of coherent population trapping (CPT) using generalised autobalanced Ramsey spectroscopy (GABRS) and combined error signal in Ramsey spectroscopy (CESRS). We consider CPT resonances excited by a coherent bichromatic field in an open Λ-system with a ‘trap’ state. Using a rigorous mathematical proof and numerical calculations, these methods are shown to lead to complete suppression of the light shift and its fluctuations. Implementation of GABRS and CESRS in CPT clocks can markedly improve accuracy and long-term stability of these devices. These methods can also be applied in atomic CPT magnetometers and interferometers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.