Abstract

The 3ω technique has become a popular method for determining the thermophysical properties of microscale and bulk materials. The prerequisite fabrication of a highly linear metal line a few hundred nanometers thick on the sample can be a failing point in specific material systems. This difficulty can be overcome by utilizing a bidirectional experimental geometry that employs a contact resistance between the sample and heating wire, which also allows for data collection under varying axial pressure loads. In this work, such a system is demonstrated with an emphasis on developing a thermal mount that will optimize sensitivity to the thermophysical parameters of interest: the sample's thermal conductivity, volumetric heat capacity, and the contact resistance between the sample and mount. A general thermal model is presented that can be simplified to analyze nearly any similar system. This model is then employed to analyze a sample in the mounting scheme described with varying applied pressures to demonstrate the general feasibility of the system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.