Abstract

ABSTRACTThe study presents the possibility to use gel permeation chromatography (GPC) with atmospheric pressure chemical ionization (APCI)–mass spectrometry (MS) for the analysis of oligomeric structures having molecular weights up to 2000 g mol−1. Polymethylmethacrylate and polystyrene oligomers were analyzed in chloroform using 2.1-mm-internal diameter GPC columns. The low flow rate of 50 µL min−1 and the ionization with chlorine allow a fast optimization of the method parameters. Excellent correlation with the expected isotope distributions and linear calibration curves were obtained for both types of oligomers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.