Abstract
AbstractImproved instrumentation facilitates on‐column injection of up to several hundreds of microliters onto GC capillary columns. It incorporates an autosampler with adjustable injection speed, a standardized precolumn system, an early vapor exit, and software that models evaporation of a given solvent at the flow and temperature conditions chosen by the user. The software calculates the appropriate injection speed, guides the autosampler correspondingly, and closes the vapor exit at a moment leaving in the precolumn the amount of unevaporated solvent that is selected by the user. © 1995 John Wiley & Sons, Inc.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.