Abstract

A baseline restorer (BLR) for pulse amplifiers is described in which current to the memory capacitor can be gated off during a pulse. This technique reduces the BLR recovery time without introducing artifacts when a short-period undershoot accompanies the input signal. The gating and reset thresholds are adjustable, as is the BLR symmetry. Tests with a pulse generator show that the baseline shift with 5V pulses at duty factors up to 75% is less than 1mV. When the signal source is a radiation detector, the gated BLR is usable at duty factors up to 90%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.