Abstract

An unusual multi-drug-resistant Pseudomonas aeruginosa (MDR-PA) was isolated in four patients whilst hospitalized in a French teaching hospital between May and August 2011. All four patients had undergone an oesophago-gastro-duodenoscopy with the same gastroscope over a five-month period. This endoscope was associated with a culture positive for the MDR-PA. Observations of endoscope reprocessing identified deviations from the agreed processes: insufficient initial cleaning, shortening of the immersion time and brushing time, insufficient channel flushing, and inadequate drying prior to storage. Since withdrawing the gastroscope and institution of strict adherence to the agreed processes, no other MDR-PA cases have been isolated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.