Abstract
The Czech Metrology Institute performs calibration and verification of noble gases stack monitors in nuclear power plants and nuclear research facilities. Together with Kr-85 and Xe-133, the radionuclide Ar-41 is measured using HPGe detectors and its activity is determined using a gamma-ray peak at 1293 keV. The counting efficiency used in these measurements was calculated by the Monte Carlo method using the MCNP code. Radioactive gas standard is prepared by irradiation of argon in a high-pressure vessel by a Cf-252 neutron generator. The inner shape and thickness of the cylinder walls were determined by radiography. The argon volume under normal conditions is determined from the high-pressure vessel volume and by a precise gas pressure measurement. As a result, the activity concentration of Ar-41 at normal conditions is certified.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.