Abstract

In the US, restrictions on the release of radioactive effluents from PET (positron emission tomography)/cyclotron facilities are typically imposed by State regulatory agencies and may be based on various methodologies and limits published by numerous agencies. This work presents suitable effluent concentration limits for various chemical forms of radioisotopes routinely produced in PET/cyclotron facilities. They were determined by application of metabolic models defined by ICRP 53 [International Commission on Radiological Protection, ICRP Publication 53 (1988)] and ICRP 26/30 which will result in compliance with effective dose equivalent limits of 100 mrem per year at the release point. The NIH Cyclotron Facility effluent air monitoring system, environmental dosimetry program, and simple, effective systems for radioactive effluent minimization are also described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.