In this procedure for disopyramide in serum, the drug is extracted into n-heptane/isobutanol (96/4 by vol), then back-extracted into 1 mol/L H2SO4. The acidic solution is made basic with sodium hydroxide, extracted with diethyl ether, and the extract evaporated. The residue is redissolved in ethanol and analyzed by gas-chromatography, with use of a nitrogen-selective detector. p-Chlorodisopyramide is used as internal standard. Concentration and instrument response for serum extracts are linearly related from 1 to 5 mg/L, the slope being 0.61, the y-intercept -0.10, the standard error of estimate 0.01, and the correlation coefficient 0.99. Within-run precision was 6 and 4% for 3 and 5 mg/L concentrations, respectively, with a between-run precision of 7% at the 3 mg/L concentration. Diazepam interferes, but procainamide, chlordiazepoxide, quinidine, lidocaine, propranolol, sulfanilamide, and many other basic drugs do not.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.