Abstract
Gas Cherenkov detectors provide a time resolved measurement of the fusion burn in inertial confinement fusion experiments. The fusion rate delivers critical benchmark figures, such as burn width and bang time. Recent detector improvements pushed temporal resolution to 10ps to make burn width measurements on igniting targets possible. First high temporal resolution measurements using CO2 gas fills had a background signal with a long decay length (tail), which was caused by gas scintillation. This gas scintillation limits the ability of the detector to resolve short burn width and high frequency features in the fusion rate measurements. A thorough investigation of the cause of the tail and mitigation options for gas scintillation is presented here. As a near-term resolution, neon gas is being used to extract fusion burn histories. Paths forward for the next generation of gas Cherenkov detectors are identified including the usage of oxygen as a Cherenkov medium.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.