Abstract
Abstract A rapid method for the analysis of dip washes is described which eliminates the usual solvent extraction procedure. The dip wash is initially diluted with acetone and then with petroleum ether. The diluted dip wash is analyzed by gasliquid chromatography, using an alkali ionization detector sensitive to phosphorus compounds. The method was applied to the determination of dioxathion (2,3-ρ-dioxanedithiol S,S-bis(O,O-diethyl phosphorodithioate)), and quintiofos (O-ethyl O-8-quinolyl phenylphosphonothioate) dip washes. Average recoveries from fouled dip washes were 100 and 104%. GLC of these compounds with an internal standard is described, which improves the precision of the method to ±2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.