Abstract

To investigate the influence of changes in thorax and lung compliance on ventilation during distal High Frequency Jet Ventilation (HFJV), similar ventilator settings were compared before and after changing the compliance in two groups of anaesthetized mongrel dogs. Each period of distal HFJV was preceded by adequate Intermittent Positive Pressure Ventilation (IPPV) to assure a start with baseline values. In Group 1 (n = 7), thorax compliance was changed by chest strapping, while in Group 2 (n = 7) lung compliance was changed by inducing acute lung injury by injection of oleic acid into the right atrium. Gas flow distribution in the ventilatory circuit during distal HFJV was determined before and after the changes in compliance were induced. Comparing similar ventilator settings during distal HFJV in the same dog, a decrease in lung or thorax compliance led to changes in gas flow distribution in the ventilatory circuit. Entrainment was decreased and bypass increased with a negative effect on the gas volume entering the lungs and on gas exchange. The results indicate that distal HFJV should be regarded as pressure-limited ventilation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.